    Description

    Dissolved oxygen sensor with an embedded EOM (Electro Optic Module) and embedded temperature and pressure sensors for monitoring of oxygen in both gaseous and aqueous environments. The outer diameter of its stainless-steel probe is 5/8" (15.875 mm) and it is 10" (25.4 cm) long.
    This fluorescence quenching based fiber optics sensor is biocompatible and can be used to monitor dissolved oxygen in biological environments, as well as dissolved oxygen monitoring in wastewater treatment. The sensor patch (zero drift-AP series) held in place by a screw-on cap and it has a
    white acrylic overcoating
    to enhance the reflected signal, with a response time of about 1 min. The sensor cap can be replaced with another cap or a fiber optic probe.
    The DOPO2 connects to a PC via USB connection and can be configured with single-channel LED excitation and detection, and modulation frequencies up to 31.2 kHz. The on-board pressure transducer and thermistor measure atmospheric pressure and sample temperature. Sensor cap is rated up to 55 bar pressure.
    Applications:
    - Waste water treatment
    - Marine Science
    - in-line measurement of oxygen in food and beverage packaging
    - life science, biological, biochemical, and biomedical process monitoring
    - Dissolved oxygen monitoring in power plants
    DOPO2 Viewer Software and standard factory calibration is included.
    DOPO2 (EOM) Specifications:
    Dimensions:
    OD: 16 mm
    L: 260 mm
    Weight:
    179 g
    Principle:
    Photoluminescence quenching using organo-metallic compounds in sol gel host
    Parameters measured:
    Gaseous oxygen partial pressure, dissolved oxygen concentration, temperature, pressure (via onboard pressure transducer)
    Compatibility:
    Gases, liquids (non-organic)
    Control Software:
    Windows 7, 8, 10 control software with data logging capability; controls include: modulation frequency, data rate, signal averaging, LED gain, LED intensity
    Sampling Rate:
    Acquires data samples every 0.1 second (10 Hz) (Software can be configured to different sample rates)
    NOTE: This is not the response speed of the probe, just the rate at which the probe is digitally sampled
    Pressure Measurement:
    On-board pressure transducer monitors atmospheric pressure
    4-20 mA Output Accuracy (at 25˚C):
    +/- 0.1%
    Electronics Operating Temperature Range:
    0 °C
    to +55 °C
    External Power Supply Voltage Range:
    4.75 to 5.25V
    Power Consumption:
    100 mA
    Communications:
    USB
    AP series Sensor Specifications
    Response rate
    1 min.
    Dynamic range in gas
    0-100% O2/0-760 mmHg O2
    Dynamic range in water (DO)
    0-40 ppm (wt)
    Dynamic range in other liquids
    0-oxygen saturation level
    Resolution in gas
    0.005% O2 / 0.04 mmHg
    Resolution in water (DO)
    0.002 ppm (wt) / 2.0 ppb (wt)
    Resolution in other liquids
    0.005% of oxygen saturation level
    Accuracy
    5% of reading
    Lowest detectable level (gas)
    0.005% O2 / 0.04 mmHg
    Lowest detectable level (water)
    0.002 ppm (wt) / 2.0 ppb (wt)
    Lowest detectable level in other liquids
    0.005% of oxygen saturation level
    Sensor drift
    Zero drift long term continuous operation
    Operating temperature
    -50 to + 80 C
    Recommended sensor cap replacement